Input & output calibration too for EuroPi.
Calibration requires the following physical setup:
- EuroPi must be connected to rack power
- For low-accuracy calibration you need either another EuroRack module or an external voltage source capable of generating precise voltages of either 5V or 10V
- For high-accuracy calibration you need another EuroRack module or adjustable external voltage source capable of generating precise voltages of 1.0, 2.0, 3.0, ..., 9.0, 10.0V.
Calibration is interactive with instructions displayed on-screen. These instructions are summarized below:
- Make sure the module is powered from your Eurorack power supply.
- Select input calibration mode by turning
K2
. PressB2
when ready a. low-accuracy with 10V input b. low-accuracy with 5V input c. high-accuracy with variable 0-10V input - Disconnect all patch cables from the module. Press
B1
- Connect your voltage source to
AIN
. PressB1
when instructed. a. low-accuracy with 10V input: connect 10V toAIN
b. low-accuracy with 5V input: connect 5V toAIN
c. high-accuracy: read the on-screen instructions and connect the specified voltages when required. - Connect
CV1
directly toAIN
. PressB1
. Wait for the module to perform the output calibration. - Repeat step 5 for
CV2
,CV3
, etc... until all CV outputs are calibrated - Reboot the module when prompted. The new calibration will be applied automatically.
Calibration data is saved to /lib/calibration_values.py
. DO NOT delete this file; if you delete it you will have to
complete the calibration again.
After calibrating and rebooting the module it is recommended to run the Diagnostic
program to verify that the
inputs and outputs have been properly calibrated.