Skip to content

Conversation

@bparks13
Copy link
Member

This PR supersedes #493.

Adding support for single-shank Neuropixels 2.0 probes, with the additional functionality of allowing the user to change the probe type in the GUI.

Note that due to the restructuring of the abstract/concrete classes, in the XML file the name of the property is ProbeConfiguration for the quad-shank probe (this is simply the name of the property, and is what all older versions of the files will be named), but for the single-shank probe it is renamed to SingleShankProbeConfiguration. We could keep the same name (ProbeConfiguration) and add the type name in the XML tag (i.e., xsi:type="onix1:SingleShank"), but this will break backward compatibility, since the deserializer cannot know which concrete type to use for the probe configuration property when loading old workflows. If we want to break compatibility for this, we can inform users that they can add the line xsi:type="onix1:QuadShank" to the ProbeConfiguration properties to let it load, but it seemed easier to maintain backward compatibility.

@anjaldoshi We do not have any single-shank probes to test with, could you build this branch and run some recordings with a single-shank probe to make sure it records data correctly?

Fixes #446

@bparks13 bparks13 added this to the 0.7.0 milestone Nov 17, 2025
@bparks13 bparks13 self-assigned this Nov 17, 2025
@bparks13 bparks13 marked this pull request as draft November 18, 2025 18:58
@bparks13
Copy link
Member Author

Before merging, make sure that the question of channel numbers is answered; there are differences in the documentation, between the given equations and the spreadsheet with the channel numbers for each electrode. We want to make sure that we have the correct channel map before adding this functionality.

- This is to match the electrode channel mapping spreadsheet, which has the correct values; the documentation contains the wrong equations, which were originally used
@bparks13 bparks13 marked this pull request as ready for review November 19, 2025 14:58
@bparks13
Copy link
Member Author

Channel mapping is now corrected, as per confirmed from Josh and Jan. The documentation contains the wrong equations, but the electrode channel mapping spreadsheet has the correct values; the equations now implemented return the correct channel number for each electrode, same as Neuropixels-PXI plugin.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment