38 changes: 38 additions & 0 deletions examples/MVLinac/TestComponent.txt
Original file line number Diff line number Diff line change
Expand Up @@ -20,4 +20,42 @@ includeFile = Flattener18MV.txt #Comment out at least one of the flatteners
#includeFile = WaterBox.txt #Comment out at least one of the dose scoring regions
#includeFile = WaterPhantom.txt #Comment out at least one of the dose scoring regions

# Source selection (electron or x-ray) and number of histories in run
ic:So/ElectronSource/NumberOfHistoriesInRun = 100
#ic:So/XraySource/NumberOfHistoriesInRun = 100

# Source at nominal position of target
# Parent is the gantry with coordinates IEC_G (see Main.txt)
s:Ge/BeamPosition/Parent = "IEC_G"
dc:Ge/BeamPosition/Pos = -1.0 cm
d:Ge/BeamPosition/TransZ = Ge/SAD - Ge/BeamPosition/Pos cm

# Electron source
s:So/ElectronSource/Type = "Beam"
s:So/ElectronSource/Component = "BeamPosition"
s:So/ElectronSource/BeamParticle = "e-"
d:So/ElectronSource/BeamEnergy = 6.51 MeV #6 MV Oncor
u:So/ElectronSource/BeamEnergySpread = 4.2 #10% FWHM 6 MV Oncor
s:So/ElectronSource/BeamPositionDistribution = "Gaussian"
s:So/ElectronSource/BeamPositionCutoffShape = "Ellipse"
d:So/ElectronSource/BeamPositionCutoffX = 1.2 mm #2 standard deviations
d:So/ElectronSource/BeamPositionCutoffY = 1.2 mm
d:So/ElectronSource/BeamPositionSpreadX = 0.6 mm #1.5 mm FWHM
d:So/ElectronSource/BeamPositionSpreadY = 0.6 mm
s:So/ElectronSource/BeamAngularDistribution = "None"

# Monoenergetic x-ray source
s:So/XraySource/Type = "Beam"
s:So/XraySource/Component = "BeamPosition"
s:So/XraySource/BeamParticle = "gamma"
d:So/XraySource/BeamEnergy = 2 MeV
u:So/XraySource/BeamEnergySpread = 0.1
s:So/XraySource/BeamPositionDistribution = "Gaussian"
s:So/XraySource/BeamPositionCutoffShape = "Ellipse"
d:So/XraySource/BeamPositionCutoffX = 1.2 mm #2 standard deviations
d:So/XraySource/BeamPositionCutoffY = 1.2 mm
d:So/XraySource/BeamPositionSpreadX = 0.6 mm #1.5 mm FWHM
d:So/XraySource/BeamPositionSpreadY = 0.6 mm
s:So/XraySource/BeamAngularDistribution = "Flat"
d:So/XraySource/BeamAngularCutoffX = 14. deg #50 cm diameter collimator
d:So/XraySource/BeamAngularCutoffY = 14. deg